Our Technology, solution for Tomorrow too!

Measuring Instrument & Tester

AC mΩ Meter

MODEL 3566

  • ■Battery inspection for measurement and comparison of battery internal resistance and open-circuit voltage
  • ■Six ranges from 30 mΩ to 3 kΩ are availeble
  • ■Fast sample rate and response speed: 60 counts/s, 84ms
  • ■Available 10KHZ low-resistance meter for fuel cell (Model 356E)

Operation manual

Typical Application

Internal resistance of Battery(Fuel Cell) , Contact resistance

Model Designation

3566 - -
   
① Data Output
Code Description
3566 Without Option
3566-03 With BCD data output (TTL level)
3566-04 With BCD data output (open collector)
3566-05 With RS-232C
3566-06 With RS-485
② Judgement Output
Code Description
Nil(blank) Open collector NPN
RY Relay output

Basic Specifications

Measurement ranges 30 mΩ, 300 mΩ, 3, 30, 300, and 3 kΩ, 6 ranges
Accuracy ± 0.5 % of rdg. +8 digit at SLOW sampling mode
± 0.5 % of rdg. +11 digit at MEDIUM sampling mode
± 0.5 % of rdg. +6 digit at FAST sampling mode
Measuring frequency AC1 kHz ± 0.2 Hz
Measurement current From 7.4 mA to 1.5 µA
Max. applied voltage 60V DC for all ranges
Sampling time 50/sec at 50HZ, 60/sec. at 60HZ (FAST mode)
6.25/sec at 50HZ, 7.52/sec. at 60HZ (MEDIUM mode)
1.56/sec at 50HZ, 1.88/sec. at 60HZ (SLOW mode)
Response speed 100 ms at 50HZ, 84 ms at 60 HZ (FAST mode)
800 ms at 50HZ, 667 ms at 60 HZ (MEDIUM mode)
1.92s at 50 HZ, 1.60 ms at 60 HZ (SLOW mode)
Measurement method AC Four-terminal measurement
Power supply 100 to 240 V AC, 50/60 HZ
Weight Approx. 1 kg
Dimensions 206 (W) x 65 (H) x 200 (D) mm
Accessories Power fuse(1), Power cord(1), Input connector (1)
Optional accessories
5803-11
probe :5811-23A
5811-21B
Kelvin clip :5811-21C

page top